PCR & extraction

Nucleic Acid Extractor System

BNP32 / BNP48

BNP32/BNP48 system is a fully-automated clinical nucleic acid extraction system that brings you walk-away automation, designed to minimize user intervention and extraction-to-extraction variability. In addition to providing quality purified eluates, the BNP32/BNP48 system offers onboard primary-sample handling and mixed-batch sampling for up to 32 or 48 samples per run.

Key features

  • Molecular-biology workflow
  • Thermal control or automated extraction
  • Configuration matched to required throughput
Made to orderPrice on request
Manufacturer
BIOBASE
Request a quote
Reply within 24 hoursValidated selection and configurationDelivery to the UK, all European countries, USA, Canada, Mexico and South America.

01 / LaboratoriX technical datasheet

LaboratoriX technical datasheet

Technical parameters consolidated from the manufacturer documentation. Each column corresponds to one model; shared values are clearly identified. Final configuration, accessories, power supply and applicable documentation are confirmed with the quotation.

ParameterBNP32BNP48
Sample Quantity3248
Processing Volume
Common to all models60µl~1000µl
Sample Volume
Common to all models20–500µl
Sample Throughput1–321–48
Magnetic Bead Recovery
Common to all models>98 %
Extracting the Difference Between Holes
Common to all modelsCV≤3 %
Heating Temperature
Common to all models8 independent heating modules, customize lysis and elution temperature(temperature range) according to your needs
Oscillating Mixing
Common to all modelsLow, medium and high third gears are adjustable, and the fluctuation range can be adjusted with the reagent volume
Reagent Type
Common to all modelsMagnetic bead open platform
Extraction Time
Common to all models8–60 min/round(depending on the reagent used)
Internal Program48 groups50000 groups
Program Management
Common to all modelsPowerful program editing capabilities to meet different reagent needs U disk program import and export can be achieved
Safety Door Design
Common to all modelsAfter the safety door is opened, the program operation will be automatically suspended, and the program can continue t run after the safety door closed
Built-in Air Duct
Common to all modelsYes
Ultraviolet Irradiation
Common to all modelsYes
Packing Size(W*D*H)580 × 510 × 700 mm700 × 520 × 750 mm
Gross Weight51 kg80 kg

MADE-TO-ORDER SUPPLY

Do you need this equipment?

Share the application, configuration, quantity and delivery location. We prepare a tailored proposal with the required documentation.